Center Point
Ideal for measuring the Young’s modulus, shear modulus, and structural damping factor of elastic and viscoelastic materials.
Description
Mecanum’s Center Point measures the Young’s modulus, shear modulus, and structural damping factor of elastic and viscoelastic materials. It works by measuring the entry mobility of a free-free beam excited at its center, a method well suited to characterizing damping treatments used to control structural noise and vibration. Paired with an optional climatic chamber, it captures how these properties change with temperature and frequency. It follows ISO 16940, SAE J3130, and JIS G 0602.
Tech specs
Measuring range
Frequency
5 to 5000 Hz
Operating temperature
0°C to 60°C
Maximum excitation force
31 N
Damping and structural loss factor, Young’s and shear modulus
No limitations, depends on beam characteristics
Dimensions
Test bench
150 (L) x 150 (W) x 200 (H) mm
Beam length range
135 to 315 mm
Maximum sample height
30 mm
Measured properties
Young’s modulus (E) or shear modulus (G)
Damping loss factor (η)
Software
ETA-X
Loss factor measurement is traditionally long and complicated. ETA-X puts decades of damping-measurement experience at your fingertips, guiding you from the beam measurement through advanced pre-processing to finished documentation in a clear workflow. Its semi-automatic analyzer mode saves significant time, and fully automated measurement lets you cover wide temperature ranges at high resolution, across up to three test benches, while you do something else. A broad set of visualization tools handles different standards and materials, a built-in nomogram generator supports ISO, WLF, and Arrhenius alpha functions, and a one-click Excel report compliant with ISO 16940, SAE J3130, and JIS G 0602 comes complete with built-in graphs.
Applications
Accessories
Optional accessories, available with your Center Point or separately.
Climatic chamber
Damping materials are viscoelastic, so their performance shifts with temperature, sometimes dramatically. Measuring at room temperature alone tells only part of the story. The optional climatic chamber lets the Center Point characterize a material across its full operating window, so you see how it behaves at the temperatures it will actually meet in service, not just on the bench.
The chamber controls both temperature and humidity around the test beam, holding stable conditions while the measurement runs. Combined with the bench’s frequency sweep, this gives you damping loss factor and modulus as functions of both temperature and frequency, the complete data set needed to model viscoelastic behavior and to match a damping solution to its application.
Key points:
• Controlled temperature across the bench’s full operating range, from 0 to 60 °C.
• Humidity control for stable, repeatable test conditions.
• Integrates directly with the Center Point test bench.
• Pilots up to 3 test benches simultaneously.
• Enables temperature-frequency characterization and 3D property surfaces.